Spectrometers - LIBS - Sensors - LEDS -
Accessories

Spectroscopy for a changing planet...  
    
Call today! 727.230.1697
info@spectrecology.com  
 
Red Tide UV/VIS Absorbance Spectrophotometer

SE-REDTIDE-UV-CUVETTE 
Ready to go UV-VIS absorbance spectrophotometer system for $2,890



The RedTide-UV-Cuvetteis a low-cost, small- footprint lab spectrophotometer that's ideal as a general-purpose instrument for budget-conscious teaching and research labs. The system consists of the following”

Red Tide Spectrometer.  This versatile low cost unit has has a wavelength range of 200-850 nm, and utilizes a detector with 650 active pixels; that’s 650 data points in one full spectrum, or one data point per nanometer. Configured with a 25 µm entrance slit, the Red Tide offers ~2.0 nm optical resolution (FWHM). The Red Tide plugs into you computer through a USB port.

The spectrometer is coupled to our  USB-ISS-UV-VIS combination sample holder and deuterium tungsten light source (200-1100 nm). This sampling system holds standard 1-cm cuvettes for easy-to-perform absorbance measurements. The USB-ISS-UV-VIS allows you to control both the intensity of the tungsten bulb and a shutter via software. The USB-ISS-UV-VIS requires an external power supply (included).

Spectrasuite Software works on Mac, Linux or Windows based computers. Spectrasuite is loaded with features: real time absorbance spectra, graphical functions for annotations and printing, time series control and strip chart mode, peak finder, automatic wavelength calibration, mapping to absorbance standards, and much more.

CVDUV1S  You get one pack of 100 disposable UV cuvettes. These cuvettes have 4 clear sides, are 1.5 – 3.0 ml volume and a1cm pathlength,

Specifications:

Physical

Spectrometer dimensions:

89.1 mm x 63.3 mm x 34.4 mm

Spectrometer weight:

190 grams

Light source dimensions:

198 x 105.1 x 40.6

Light source weight:

200 g

Detector Specifications

Detector:

Sony ILX-511 Linear CCD Detector

Pixels:

650 pixels

Pixel size:

14 µm x 200 µm

Pixel well depth:

62,500 electrons

Signal-to-noise ratio:

250:1 (at full signal)

A/D resolution:

12 bit

Corrected linearity:

>99.8%

Sensitivity:

75 counts at 400 NM; 41 photons/count at 600 nm

Optical Bench

Design:

f/4, Asymmetrical crossed Czerny-Turner

Focal length:

42 mm input; 68 mm output

Entrance aperture:

25 µm wide slit

Grating:

Grating #1

OFLV filter:

OFLV-200-850

UV enhanced window:

Yes, UV4

Fiber optic connector:

SMA 905 to 0.22 numerical aperture single-strand optical fiber

Spectroscopic

Wavelength range:

200 to 880 nm

Optical resolution:

2.0 nm FWHM

Signal-to-noise ratio:

300:1 (at full signal)

A/D resolution:

12 bit

Dark noise:

3.2 RMS

Integration time:

3ms to 65 seconds

Dynamic range:

2 x 108 (system), 1300:1 for a single acquisition

Stray light:

<0.05% at 600 nm; 0.10% at 435 nm

Light Source/Sample Holder

Power consumption:

1.8 A @ 5 VDC; external power supply (included)

Pathlength:

1 cm

Cuvette shape:

Square

Light source:

Deuterium tungsten

"Z" dimension:

15 mm

Bulb life (hours):

800 deuterium; 2,000 tungsten

Time to stabilize:

~30 minutes

Electronics

Power consumption:

250 mA @ 5 VDC

Data transfer speed:

Full scans to memory every 4 ms with USB 2.0 port, 18 ms with USB 1.1 port

Inputs/Outputs:

Yes, 8 onboard digital user-programmable GPIOs

Analog channels:

No

Auto nulling:

Yes

Breakout box compatible:

Yes, HR4-BREAKOUT

Trigger modes:

5 modes

Strobe functions:

Yes

Connector:

22-pin connector

Computer

Operating systems:

Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port

Computer interfaces:

USB 2.0 @ 480 Mbps (USB1.1 compatible); RS-232 (2-wire) @ 115.2 K baud

Peripheral interfaces:

SPI (3-wire); I2C inter-integrated circuit