Wavelength Calibration Source
Spectrometer wavelength calibration sources are available for convenient, reliable spectrometer calibration across UV-NIR wavelengths.
Options include mercury-argon (253-1700 nm), krypton (427-893 nm), neon (540-754 nm), argon (696-1704 nm) and xenon (916-1984 nm) gas-discharge emission sources. With multiple wavelength options and emission lines to utilize, users can more readily choose a source, or combination of sources, to match analytical wavelengths of interest within the measurement range.
- Gas-discharge sources — produce atomic emission lines for performing reliable spectrometer wavelength calibration
- UV-NIR coverage — choose from multiple sources across a broad range of wavelengths
- Remote on/off operation -- helps you avoid disturbing experiment setups
- Convenient power features — sources include 5 VDC power supply and embedded, rechargeable lithium ion battery
AR-2
Engineering:
Wavelength Range: 696nm – 1.704μm
Source: Argon
Amplitude Stabilization: ~1 minute
Source Lifetime: 3,500 hours (typical)
Physical:
Fiber Connector: SMA 905
Aperture: 3 mm
Dimensions: 16.7 x 10.6 x 3.4 cm (without feet)
Power Consumption: 2A (maximum) at 5 VDC when battery is charging
Weight: 180 g
Power Requirements: 5 VDC power supply; rechargeable, embedded lithium ion battery
HG-2
Engineering:
Wavelength Range: 253nm – 1.7μm
Source: Mercury Argon
Amplitude Stabilization: ~1 minute
Physical:
Fiber Connector: SMA 905
Aperture: 3 mm
Dimensions: 16.7 x 10.6 x 3.4 cm (without feet)
Power Consumption: 2A (maximum) at 5 VDC when battery is charging
Weight: 180 g
Power Requirements: 5 VDC power supply; rechargeable, embedded lithium ion battery
KR-2
Engineering:
Wavelength Range: 427nm – 893nm
Source: Krypton
Amplitude Stabilization: ~1 minute
Source Lifetime: 3,500 hours (typical)
Physical:
Fiber Connector: SMA 905
Aperture: 3 mm
Dimensions: 16.7 x 10.6 x 3.4 cm (without feet)
Power Consumption: 2A (maximum) at 5 VDC when battery is charging
Weight: 180 g
Power Requirements: 5 VDC power supply; rechargeable, embedded lithium ion battery
NE-2
Engineering:
Wavelength Range: 540nm – 754nm
Amplitude Stabilization: ~1 minute
Source Lifetime: 3,500 hours (typical)
Physical:
Fiber Connector: SMA 905
Aperture: 3 mm
Dimensions: 16.7 x 10.6 x 3.4 cm (without feet)
Power Consumption: 2A (maximum) at 5 VDC when battery is charging
Weight: 180 g
Power Requirements: 5 VDC power supply; rechargeable, embedded lithium ion battery
XE-2
Engineering:
Wavelength Range: 916nm – 1.984μm
Source: Xenon
Amplitude Stabilization: ~1 minute
Source Lifetime: 3,500 hours (typical)
Physical:
Fiber Connector: SMA 905
Aperture: 3 mm
Dimensions: 16.7 x 10.6 x 3.4 cm (without feet)
Power Consumption: 2A (maximum) at 5 VDC when battery is charging
Weight: 180 g