UV-VIS Radiometric Calibration Source
Radiometrically calibrated light sources from Spectrecology are used to calibrate the absolute spectral response of a spectrometer system. Using these light sources and our software, you can determine absolute intensity values across UV, Visible, and NIR wavelengths.
Radiometrically calibrated light sources are carefully characterized to deliver a known quantity of light with very low uncertainty. Each source is measured to NIST-traceable standards, creating an unbroken chain of traceability and providing you with the highest quality and most reliable data.
- Stable sources - provide consistent output for the most reliable data
- Multiple options - radiometrically calibrated sources for UV-Visible-NIR and Visible-NIR wavelengths
- Digital calibration data - can be loaded into software for absolute intensity data
- Applications flexibility - measure LEDs, digital displays, and other radiant sources
DH-3P-BAL-CAL
Engineering:
Wavelength Range: 230nm – 1.1μm
Nominal Bulb Power: 25 W (deuterium), 20 W (tungsten halogen)
Calibrated For: Absolute irradiance (µW/cm2/nm)
Calibration Lifetime: 50 hours
Drift (Voltage): <0.01% per hour
Minimum Uncertainty Level: 4.0% at 700nm
Stability (Current): <5 x 10-6 peak-to-peak (0.1-10.0 Hz)
Warm-up Time: 40 minutes (deuterium), 20 minutes (tungsten halogen)
Environmental:
Safety & Regulatory: CE, ROHS, WEEE
Operating Humidity Specification: 5-95% without condensation at 40°C
Operating Temperature: 5 °C – 35 °C
Physical:
Shutter: Yes
Fiber Connector: SMA 905; CC-3-UV-S or 6.35 mm barrel for cosine corrector
Dimensions: 15 x 13.5 x 28.5 cm
Power Consumption: Approximately 78VA
Weight: 6 kg
Power Requirements: 85-264 V, 50/60 Hz
DH-3P-CAL
Engineering:
Wavelength Range: 230nm – 1.1μm
Source: Deuterium & Tungsten Halogen
Nominal Bulb Power: 25 W (deuterium), 20 W (tungsten halogen)
Calibrated For: Absolute irradiance (µW/cm2/nm)
Calibration Lifetime: 50 hours
Drift (Voltage): <0.01% per hour
Minimum Uncertainty Level: 3.7% at 700nm
Stability (Current): <5 x 10-6 peak-to-peak (0.1-10.0 Hz)
Warm-up Time: 40 minutes (deuterium)
Environmental:
Safety & Regulatory: CE, ROHS, WEEE
Operating Humidity Specification: 5-95% without condensation at 40°C
Operating Temperature: 5 °C – 35 °C
Physical:
Shutter: Yes
Fiber Connector: SMA 905; CC-3-UV-S or 6.35 mm barrel for cosine corrector
Dimensions: 15 x 13.5 x 28.5 cm
Power Consumption: Approximately 78VA
Weight: 6 kg
Power Requirements: 85-264 V, 50/60 Hz
HL-3P-CAL
The HL-3P-CAL is radiometrically calibrated for use with a cosine corrector (CC-3), which is included. Also, the HL-3P-CAL can be calibrated for use with an optical fiber, which requires the purchase of the fiber and our calibration service
Engineering:
Wavelength Range: 350nm – 1.1μm
Nominal Bulb Power: 5 W
Calibrated For: Absolute irradiance (µW/cm2/nm)
Calibration Lifetime: 50 hours
Drift (Voltage): <0.3% per hour
Minimum Uncertainty Level: 3%
Stability (Current): 0.5% peak-to-peak
Warm-up Time: 15 minutes
Environmental:
Safety & Regulatory: CE, ROHS, WEEE
Operating Humidity Specification: 5-95% without condensation at 40°C
Operating Temperature: 5 °C – 35 °C
Physical:
Shutter: Yes
Fiber Connector: SMA 905; CC-3
Dimensions: 14 x 6 x 6 cm
Power Consumption: 1.2 A @ 12 VDC
Weight: 0.5 kg
Power Requirements: 85-264 V, 50/60 Hz
HL-3P-INT-CAL
Engineering:
Wavelength Range: 350nm – 1.1μm
Nominal Bulb Power: 5 W
Source: Tungsten halogen
Calibrated For: Absolute irradiance (µW/cm2/nm)
Calibration Lifetime: 50 hours
Drift (Voltage): <0.3% per hour
Minimum Uncertainty Level: 7%
Stability (Current): 0.5% peak-to-peak
Warm-up Time: 15 minutes
Environmental:
Safety & Regulatory: CE, ROHS, WEEE
Operating Humidity Specification: 5-95% without condensation at 40°C
Operating Temperature: 5 °C – 35 °C
Physical:
Fiber Connector: Integrating sphere
Dimensions: 14 x 6 x 6 cm
Power Consumption: 1.2 A @ 12 VDC
Weight: 0.5 kg
Power Requirements: 85-264 V, 50/60 Hz